Fix CS16 capture, CLI tools while connected, sample-rate range, stdout banner
- CS16 streams read into an int16 buffer; files hold raw SC16 Q11 (full scale 2048) instead of all zeros - bladeRF-cli backed tools release the USB claim while they run and restore RX/TX settings afterwards; loopback uses the plugin setting - calibrate_dc runs "calibrate lms" + "calibrate dc rxtx"; a bare "cal dc" only printed help and reported success - report the span of all SoapySDR ranges (rate was capped at 10 Msps) - parse "Firmware version" from bladeRF-cli 1.10 - startup banner goes to stderr, off the stdio JSON-RPC channel - shared RX chunk reader: settle discard, overflow and stall handling, stream always closed; captures report overflows, clipping, peak dBFS - spectrum power normalized to dBFS - unit tests, README troubleshooting, version 2026.10.25
This commit is contained in:
+1
-1
@@ -1,6 +1,6 @@
|
||||
[project]
|
||||
name = "mcbladerf"
|
||||
version = "2026.02.08"
|
||||
version = "2026.10.25"
|
||||
description = "MCP server for Nuand BladeRF SDR control via SoapySDR"
|
||||
authors = [{name = "Ryan Malloy", email = "ryan@supported.systems"}]
|
||||
readme = "README.md"
|
||||
|
||||
Reference in New Issue
Block a user